How to Retest a False Break Specimen?
False breaks can happen for many reasons. Your Forney machine provides you the ability to retest the specimen by following this guide.
- Verify that a false break of the test specimen occurred by reviewing the break type.
- On the Break Type screen select No Break -OR- Invalid to discard results.
- At the prompt asking if you want to save your data, select "No."
- Rerun the specimen test.
- If the test was successful, select the correct Break Type to record results to ForneyVault.